CardioPart 12 USB / USB-P details
Accessories / Components
Accessories
- Patient cable
Accessories
- Patient cable
Technical Data
Dynamic range | +/- 316 mV DC |
Sample rate | 8000 Hz [125 μs] for each of the 10 electrode channels |
Resolution | 1 μV/LSB [0.01 mm] |
Frequency range | 0 – 150 Hz |
Pacemaker detection | Digital monitoring of all electrodes |
Input impedance | > 50 MOhm |
Electrodes check | Frequency analysis and impedance measurement |
Input protection | Against defibrillator shock1) and HF from surgery devices |
Patient cable connection | 15-pin D-Sub for 10-lead patient cable |
Applied part | Type CF |
PC interface | USB 2.0 (5 meter USB cable) |
Power supply | Via the USB port of the PC |
R wage trigger output | Via LPT port of the PC |
Dimensions | 95 mm x 64 mm x 28 mm |
Weight | 90 g |
Standards | DIN EN 60601-1; DIN EN 60601-2; DIN EN 60601-2-25; ANSI/AAMI EC 11 |
1) The patient cable or suction electrode system must feature a protective resistor of 10 kOhm in each of the cables. Protection against defibrillator
discharge is only ensured through the use of a patient cable or suction electrode system of this type.
1) The patient cable or suction electrode system must feature a protective resistor of 10 kOhm in each of the cables. Protection against defibrillator
discharge is only ensured through the use of a patient cable or suction electrode system of this type.
Connection
Supported HIS, PACS and BDT/GDT communication | CardioPart 12 USB | CardioPart 12 USB-P (specifically for medical practices) |
---|---|---|
BDT/GDT | ||
HL7/DICOM | (Option) | — |
Options
Available ECG options | CardioPart 12 USB | CardioPart 12 USB-P (specifically for medical practices) |
---|---|---|
m Resting ECG with automatic measurement |
— | |
mr Resting ECG with automatic measurement and rhythm ECG option |
||
i Resting ECG with automatic measurement, interpretation and rhythm ECG option |
||
s Resting and stress test ECG with measurement, interpretation and rhythm ECG option, ST measurement, control of ergometers and blood pressure meters |
||
as Resting and stress test ECG with measurement, interpretation and rhythm ECG option, ST measurement, arrhythmia analysis, control of ergometers and blood pressure meters |
— | |
mrs Resting and stress test ECG with measurement, ST measurement, ergometer control and blood pressure measurement |
— |
Product Inquiry
Would you like information about our products? Simply fill out the inquiry form - we will get back to you as soon as possible.
All fields with a "*" are mandatory!